| Defibrillation |
| Defib common |
|
| Waveform |
Truncated Exponential Biphasic (e*-cube) |
| Charge Time |
Adapter: Less than 5 seconds to 200 Joules Battery: Less than 7 seconds to 200 Joules |
| AED Mode |
|
| Output Energy |
200J |
| Shock Delivery |
Via multifunction defib electrode pads |
| AED Develop guidelines |
Shock advisory sensitivity and specificity meet AAMI DF-80 guidelines |
| Voice & Text Prompts |
Guide the user through the protocol via multifunction defib electrode pads |
| Manual Mode |
|
| Output Energy (Selected) |
1~10J, 15J, 20J, 30J, 50J, 70J, 100J, 120J, 150J, 170J, 200J |
| Shock Delivery |
External paddle (with Pediatric) Internal paddle |
| Synchronous Cardioversion |
|
| Noninvasive Pacing |
|
| Waveform |
Monophasic Truncated Exponential |
| Mode |
Demand and Fixed Mode |
| Amplitude Accuracy |
0 ~ 200mA (± 5mA) |
| Pulse Width |
20ms (± 1ms) |
| Pulse Rate |
30 ~180 ppm (± 1.5%) |
| Refractory Period |
340 msec (30 to 80 ppm) 240 msec (90 to 180 ppm) |
| ECG Monitoring |
|
| Input |
3-Lead Cable: I, II, III 5-Lead Cable: I, II, III, aVR, aVL, aVF, or V |
| Lead Fault |
"Lead Fault" message and dashed line displayed, if an electrode or lead wire becomes disconnected |
| Heart Rate Display |
15 to 300 bpm (± 3bpm) |
| ECG Size |
5, 10, 20, 40 mm/mV and Auto-gain |
| Heart Rate/Arrhythmia Alarm |
HR, Asystole, VFIB, VTACH |
| 12-Lead ECG |
|
| Input: 10-Lead Cable: |
I, II, III, aVR, aVL, aVF, VI, V2, V3, V4, V5, V6 |
| Display View |
All 12-lead ECG waves display simultaneously |
| Transmission |
SD data card |
| Sp02 Pulse Oximetry |
|
| Saturation |
70-100% (± 3digits) |
| Pulse Rate |
20-250 bpm (± 3bpm) |
| Perfusion |
0.2% |
| Module Manufacturer |
Nellcor MP 100 |
| Sp02 Alarm |
Less than Minimum setting rate Over than Maximum setting rate |
| NIBP |
|
| Patient Population |
Adult, Pediatric, Neonate |
| Method |
Oscillometric |
| Control |
Automatic and manual measurements |
| Auto Intervals |
1, 3, 5, 10, 15, 30, 60, 120 min |
| Displayed Pressures |
Systolic, Diastolic, Mean mmHg |
| Displayer Units |
Adult: 40 to 260 mmHg |
| |
Pediatric: 40 to 160 mmHg |
| |
Neonate: 40 to 130 mmHg |
| Systolic Range |
Adult: 20 to 200 mmHg |
| |
Pediatric: 20 to 120 mmHg |
| |
Neonate: 20 to 100 mmHg |
| Diastolic Range |
±3mmHg |
| Pressure Transducer |
Adult: 300 mmHg Accuracy |
| |
Pediatric: 300 mmHg |
| |
Neonate: 150 mmHg |
| Redundant Circuit Overpressure Limit |
| Date Storage |
|
| External memory card |
SD Card (ECG data, Event, Voice) |
| Printer |
|
| Continuous ECG Strip |
Real-Time (8 Seconds delay) |
| Auto Printing |
Recorder can be configured to print marked events, charge, shock and alarms |
| Printing Speed |
25mm/s |
| Paper |
50mm Width / 40mm Diameters |
| Automatic Self-Test |
Power on Self-Test |
| |
Run Time Self-Test |
| |
Manual Self-Test |
| |
Periodic Self-Test (Daily / Weekly / Monthly) |
| Power Source |
|
| External Battery Pack |
Lithium Polymer |
| Type |
14.8V 3.1Ah (Rechargeable) |
| Capacity |
When new, minimum of 100 shock deliveries (200J) |
| AC Power Pack |
|
| Output |
18V, 6A
|
| Display |
|
| LCD Dimensions |
7 inch Diagonal (152mm * 91mm) |
| Type |
TFT Color |
| Resolution |
800 * 480 pixels |
| Wave Viewing Time |
5 seconds (ECG) |
| Dimensions |
|
| Size |
318 x 208 x 355 (W x L x H, mm] |
| Weight |
5kg [with external paddle] |
| Parts & Accessories |
|
| Standard Package |
Device |
| |
External Paddle (Adult, Pediatric) |
| |
3 Lead ECG cable |
| |
ECG electrodes |
| |
Defi pads & Adaptor |
| |
Built in printer -Power cord &SMPS |
| |
Internal Battery -Gel |
| |
User's Guide |
| Optional |
Carrying case |
| |
SD card |
| |
SD card reader |
| |
Pediatric pads |
| |
5 Lead ECG cable |
| |
Car cigar Lighter jack |
| |
Sp02 Module Set |
| |
Pacer |
| |
10 Lead ECG cable |
| |
NIBP (Non-Invasive Blood Pressure) |
| |
Software for data management |