Measures: EtCO2 with Capnograph, Respiration Rate, Spo2 with Plethysmogram, Pulse Rate & Printer.
EtCo2 Monitor is an extremely useful and comprehensive anesthetic monitoring device; it can supervise Four Physical parameters at the same time namely Density of EtC02, Respiration, Pulse Rate and Oxygen Saturation. It has a crisp clear wave form with large bright numerical readouts for maximum visibility. It has a user selectable option to view Capnograph or Plethysmogram Waveform simultaneously.
ECG
- Lead mode : 3 - lead or 5 - lead
- ECG waveform: 2 channals
- S-T segment detection: Measurement renge : -2.0mv~2.0mv
- Alarm: Yes, audible and visual alarm, alarm events review
- Bulit-in rechargeable lithium battery
- 72-hours graphic and tabular trends of all parameters
- 32 seconds full-disclosure waveform review
- 500 NIBP measurement date can be storage and recall
- Date and waveforms color be adjustable
- Suitable for adult, pediatric and neonatal patient
- Lead selection : I; II; III; avR ; avL; avF; V
- Accuracy: +/-lbpm or +/-1 %, which is greater
- ECG waveforms of 7-leads display on the same screen
- 72 alarm events of all parameters recall 32 seconds
- Arrhythmia analysis and S-T segment anyalysis
| End Tidal CO2 (EtCO2) with Capnograph |
| Measurement Range |
0% - 10% CO2 |
| Units (User selectable) |
mmHg or % |
| Display Range |
0 - 100mmHg or 0 - 10.0% |
| Accuracy |
±1% when density of EtCO2 in 0.0 - 6% |
| |
±2% when density of EtCO2 in 6 - 10% |
| Technique |
Non Dispersive Infra Red Absorption with Automatic Offset Calibration Technology and moisture removal system |
| Respiration Rate (RR) |
| Measurement Range |
0-150bpm |
| Range |
6times/min - 60times/min |
| Deviation |
The bigger value of 2% or 2times/min |
| Averaging |
4 breath average |
| Oximetry Saturation (SpO2) With Plethysmogram (Pleth Wave) |
| Range |
0% - 99% |
| Accuracy |
when 50% - 79%, ±3% (absolute Value) |
| |
when 80% - 99%, ±2% (absolute value) |
| |
Reads through Motion and Low Perfusion Condition. |
| Perfusion |
Compatible for Adult, Pediatric and Neonate Sensors. |
| Pulse Rate (PR) |
| Range |
30bpm - 250bpm |
| Accuracy Deviation |
when 30bpm - 40bpm, ±2bpm |
| |
when 41bpm - 250bpm, ±5bpm |
| Alarm |
User Adjustable Upper and lower Limits for all EtCO2, Resp., SpO2 & Pulse Rate with audio & Visual Alarm. |
| Apnea Alarm |
Yes |
| Accuracy |
0.2 % |
| Trend |
24 Hours Trend for all Parameter: Etco2 Concentration, Respiration Rate, Spo2 & Pulse. |
| Printer |
50mm Real time/Trend EtCO2 & SpO2 with Graph |
| Power & Working Requirement |
| Temperature |
10°C to 60°C |
| Voltage |
AC 200V- 240V |
| Power Consumption |
100W |
| Battery Backup |
Two Hour |
|
|
 |
| |